Apogee Instruments SU-300 UV-B Manuale d'uso - Pagina 17

Sfoglia online o scarica il pdf Manuale d'uso per Accessories Apogee Instruments SU-300 UV-B. Apogee Instruments SU-300 UV-B 19. Ultraviolet index & ultraviolet-b sensor

Apogee Instruments SU-300 UV-B Manuale d'uso

TROUBLESHOOTING AND CUSTOMER SUPPORT

Independent Verification of Functionality
Apogee SU-300 sensors are self-powered devices and output a voltage signal proportional to incident UV-I/UV-B
radiation. A quick and easy check of sensor functionality can be determined using a voltmeter with millivolt
resolution. Connect the positive lead of the voltmeter to the white wire from the sensor and the negative lead (or
common) to the black wire from the sensor. Direct the sensor head toward the sun and verify the sensor provides
a signal. Blocking all UV radiation from the sensor should force the sensor signal to zero.
Compatible Measurement Devices (Dataloggers/Controllers/Meters)
SU-300 UV-I/UV-B sensors are calibrated with a standard calibration factor of 10.0 UV Index units per mV, yielding
a sensitivity of 0.1 mV per UV Index unit. Thus, a compatible measurement device (e.g., datalogger or controller)
should have resolution of at least 0.1 mV to provide a measurement resolution of 1 UV Index, and 0.01 mV to
provide a measurement resolution of 0.1 UV Index.
An example datalogger program for Campbell Scientific dataloggers can be found on the Apogee webpage at
http://www.apogeeinstruments.com/content/UV-Sensor.CR1.
Zero Offset Error
With the use of certain dataloggers it is possible to measure a non-zero voltage (zero offset) when the sensor
output should be zero (no UV irradiance incident on diffuser). This offset can be corrected by adding or subtracting
the measured offset from the sensor output. However, if the offset is substantial, and your sensor is outputting
unrealistic values then it may need to be recalibrated. In this case, contact Apogee customer support to recalibrate
the sensor.
Cable Length
When the sensor is connected to a measurement device with high input impedance, sensor output signals are not
changed by shortening the cable or splicing on additional cable in the field. Tests have shown that if the input
impedance of the measurements device is greater than 1 mega-ohm there is negligible effect on the calibration,
even after adding up to 100 m of cable. All Apogee sensors use shielded, twisted pair cable to minimize
electromagnetic interference. For best measurements, the shield wire must be connected to an earth ground. This
is particularly important when using the sensor with long lead lengths in electromagnetically noisy environments.
Modifying Cable Length
See Apogee webpage for details on how to extend sensor cable
length at
http://www.apogeeinstruments.com/how-to-make-a-
weatherproof-cable-splice/.
NOTE: Remove protective cap before using any sensor.