DUTSCHER MagnetaPure 32 Plus Manuale di avvio rapido - Pagina 2
Sfoglia online o scarica il pdf Manuale di avvio rapido per Apparecchiature di laboratorio DUTSCHER MagnetaPure 32 Plus. DUTSCHER MagnetaPure 32 Plus 2. Nucleic acid purification system
MagnetaPure 32 Plus Nucleic Acid Purification System
3. Plate installation
Prepare the plates according to the manufacturer's protocol instructions. Open the cabin door and place
the appropriately prepared 96-well plates with the sample in the locating slot according to the arrow. Pay
attention to the orientation of the plate: Place the A1 well of the plate so that it is in the upper left
corner.
96-Deep well plate
4. Install the magnetic rod cover
Insert the tip combs into the slots on the mounting groove of the tip comb holder according to the arrow.
Ensure the tip combs are fully insertes as it may damage the magneti rods if installation is incorrect.
Never operate the MagnetaPure 32 Plus without the mangetic rods being covered with the tip combs.
Magnetic rod's tip
5. Operation
For details on operation, setting and operation, please refer to the operating instructions.
2