DV Power TRT03 Manuale - Pagina 6
Sfoglia online o scarica il pdf Manuale per Apparecchiature di prova DV Power TRT03. DV Power TRT03 19. Three-phase transformer turns-ratio test instrument
TRT03
2 Description
2.1 Front Panel Components
1. Mains Power Connector and Power Switch
Mains power connector
Connect TRT03 to the mains power supply with a power cord.
Power switch – Double pole switch
•
I In this position, TRT03 is connected to the mains power supply.
•
0 In this position, TRT03 is separated from the mains power supply.
2. Protective Earth Connector
For protection against parasitic currents or voltages, always connect TRT03 protective earth
connector to the protective ground (PE). Use only the original cable.
For safety reasons, always establish this connection before establishing any other connection, and
remove this connection as the very last step.
3. Operator Control
Display
Displays the settings during the device programming as well as the measured values during a test
operation.
M-T003NN-205-EN
6
Description