ADE M501020 Manuale operativo - Pagina 11

Sfoglia online o scarica il pdf Manuale operativo per Bilance ADE M501020. ADE M501020 20. Electronic wheelchair scales

ADE M501020 Manuale operativo
 Configuration mode (set-up mode)
The values for different parameters can be set in configuration mode and the display unit calibrated.
This function is reserved for service personnel.
 Test mode
There are different diagnostic functions for the ADE-MAK in test mode; this is reserved for service
personnel.
Manipulations to test and configuration modes can revoke the calibration status and
will result in forfeiture of any warranty claims!

Settings

The following functions are available in setting mode:
Select the serial interface, no metrological parameters.
Set date, time and consecutive printout numbers.
Set print format.
Set switching points.
Test functions
Access the function menu by pressing the CLEAR (13) key until TIME appears on the display. Use the
key pad (13) to navigate within the menus, to select numerical values and to increase or lower the
values. Press the ENTER key to save the values entered and to return to the super ordinate menu
level.
Menus structure of setting mode:

Setting date and time

Hold the CLEAR key pressed for approx. 4 sec; FILTER appears on the display.
Press key "6" 6x; TIME appears on the display.
Press ENTER key; SHOW appears on the display.
Press key "6" 1x; HOUR appears on the display.
Press ENTER key and set the hours using the arrow keys "2", "4", "6" and "8".
Press the ENTER key and set the minutes and seconds.
Go back by pressing key "2"; TIME appears on the display.
Press key "6" 1x; DATE appears on the display like the time entry.
Press key "2" 2x; EXIT Y appears on the display.
Confirm with ENTER; date and time are set.

Aligning the scales

Place the scales onto a solid, even surface
Align the scales so that all feet are in contact with the surface.
Check that the scales are horizontal using a spirit level ensuring that the air bubble is centred.
Connect the display unit to the scales platform using the connecting plug provided.
If the scales have been transported or were in storage, they will need to be re-aligned as described
above.
M501020-210322-Rev006-UM-e

Operating the scales

11