Error Message and Trouble Shooting
Message
Cause
• Problem with code strip.
• Insert the strip in an
improper way.
• Problem with test strip.
• The power of the battery
is too low to run a test.
• Problem with code strip.
• Insert the code strip in
an improper way.
• The temperature was
below or above the
meter operating range.
• An used test strip or a
damp strip.
23
Solution
Repeat the coding procedure and ensure the code
number is the same in label, code strip and
screen. Insert the strip again. If the problem
persists, please contact the distributor for service.
Replace a new battery immediately.
Repeat the coding procedure and ensure the code
number is the same in label, code strip and
screen.
If the problem persists, please contact
the distributor for service.
Repeat the test after the meter has reached to the
temperature within the operating range. If the
problem persists, please contact the distributor
for service.
Review the instructions and try again with a new
test strip.
If the problem persists, please contact the
distributor for service.
• Uncompleted test caused by
Review the instructions and try again with a
removing the test strip during
new test strip. Please don't remove the strip
the measurement.
before the test is completed.
• Problem with strip.
Review the instructions and try again with
the same test strip.
If the problem persists, please contact the
distributor for service.
• Use an improper code
Repeat the coding procedure and ensure the
code number is the same in label, code strip
strip.
and screen.
contact the distributor for service.
• Do not get enough
Take a new strip and make sure you have
sample volume for the
enough sample volume for your test. Repeat
the test again. If the problem persists,
test.
please contact the distributor for service.
• Test result is higher than
1.Review the instructions and try again with
the range listed on Chapter 9.
a new test strip.
2.If the problem persists, please call local
authorized distributor for help.
• Test result is lower than
1.Review the instructions and try again with
the range listed on Chapter 9.
a new test strip.
2.If the problem persists, please call local
authorized distributor for help.
If the problem persists, please