10. EMC Related Notification
Warning: Medical electrical equipment needs special precautions regarding EMC and needs to be
installed according to the EMC information provided. Careful consideration of this information is
essential when stacking or collocating equipment and when routing cables and accessories.
Warning: RF mobile communications equipment can affect medical electrical equipment.
The Universal Therapy System (UTS) is intended for use in the electromagnetic
environment (for home and professional healthcare) specified below.
The customer or the user of the Universal Therapy System (UTS) should assure that it is
used in such an environment.
Emission test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations
/flicker emissions IEC
61000-3-3
During test, the wired control box signal display flashed but could return to normal status after the test
was not affected so there is no concern with the basic safety.
Please refer to Risk management Report for the detail.
Manufacturer's declaration-electromagnetic emissions
Compliance
Group 1
Class B
Not applicable (120V)
Class A (230V)
Not applicable (120V)
Compliance (230V)
Electromagnetic environment-
(for home and professional
healthcare environment)
The Universal Therapy System (UTS)
uses RF energy only for its internal
function. Therefore, its RF emissions
are very low and are not likely to
cause any interference in nearby
electronic equipment.
The Universal Therapy System (UTS)
is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.
11
guidance