Great Life Healthcare UTS Manuale d'uso - Pagina 4
Sfoglia online o scarica il pdf Manuale d'uso per Apparecchiature mediche Great Life Healthcare UTS. Great Life Healthcare UTS 18.
2. The Purpose of this Manual
This operation manual is mainly focused on the set-up, cleaning and routine maintenance of the
Universal Therapy System (UTS). We recommend that you keep this manual in a safe location to help
answer any questions that may arise relating to the system.
3. Product Description – Intended Use
The Universal Therapy System (UTS) is an alternating air replacement system designed to prevent and
treat pressure ulcers. The Universal Therapy System (UTS) works on the principle of actively
encouraging tissue blood flow by gently increasing and decreasing pressure within the support surface
over a 10-minute cycle. This results in the reduction of surface pressures helping to prevent tissue
breakdown and encourage healing.
Contraindications: Patient conditions for which the application of pressure therapy on the Universal
Therapy System (UTS) systems. Contraindications include:
*Instable spinal cord injury
*Cervical traction
Intended Care Setting will be in Aged care and Home care.
Intended User Profile:
●
At least fifteen years old
●
Language understanding:
-
Read and understand 'Westernized Arabic' numerals when written in Arial font
Understands hygiene
-
●
Experience Requirement: Have nursing facilities
●
Permissible impairments: Except for contraindications
For added durability, and longer working life, the Universal Therapy System (UTS) has been constructed
using high strength components and premium grade materials. The mattress features a heavy-duty
base layer and a water resistant, multi-stretch vapor permeable PU cover to reduce risks of friction and
shear.
4. General Safety
It is important to read the information in this user manual before you use your Universal Therapy
System (UTS).
Please follow the guidelines below for your safety and maintaining of system performance.
●
Maximum patient weight is 350 lbs. and minimum user weight is 66 LBS.
●
Avoid exposing pump to liquids.
●
When cleaning do not use of Phenol based substances.
●
This Universal Therapy System (UTS) must be used on top of a bed frame