Emergency Caller Products LT-900 Manuale d'uso

Sfoglia online o scarica il pdf Manuale d'uso per Trasmettitore Emergency Caller Products LT-900. Emergency Caller Products LT-900 2. Universal

Emergency Caller Products LT-900 Manuale d'uso
1) The Universal Transmitter LT-900 with
pre-installed transmitter code.
2) Wrist Strap (not shown)
LT-900 Long Range Transmitter is Long Life
battery operated and connects to Sensor Pads.
The transmitter is attached to posts on
Wheelchair, Beds, etc. with Fast Install Velcro
straps. When sensor is activated by weight of
patient (getting ON or OFF sensor pad), the
transmitter sends the signal to the Enunciator.
ON-OFF switch allows caretaker to attend
patient. The LT-900 transmitter is selectable to
monitor chair, bed, floor and rail/bed sensor
pads. Transmitter is selectable for INSTANT or
DELAY (2 sec.) operation, allowing patient to
move around in chair or bed without sending
alarm!
The LT-900 Long Range Transmitter has a
switch to turn the transmitter on for instant or
delayed operation or off to allow the caregiver
to work with a patient and not have the alarm
sound.
The led will flash when transmitter is activated.
In an alarm condition make sure both transmit-
ter and receiver units are reset to there non-
alarm condition when situation has returned to
normal.
There is also a switch to set the sensor for
Normally Open, position O, to activate the
alarm when weight is getting on the pad or
Normally Closed, position C, to activate the
sensor when weight is getting off the pad.
We strongly recommend that you
test the transmitter with the receiv-
er/enunciator system frequently.
From the time of installation, it is
absolutely necessary to test the
system at least once a week to
ensure that the LT-900 (transmit-
ter) will be ready when needed.