- Low battery voltage fault.
- High battery voltage fault.
- Left motor (or connection) fault.
- Right motor (or connection) fault.
- Left or right park brake (or connection) fault. ‧‧‧‧‧-‧‧‧‧‧-
- Controller fault.
- Motor stalled or joystick out of neutral time out. ‧‧‧‧‧‧‧-‧‧‧‧‧‧‧-
"‧" means all 6 LED's flash on for a period of 0.2seconds and off for 0.8 seconds.
" -" means all 6 LED's off for a period of 2 seconds before flashing sequence repeats.
If more than one fault exist then the fault having the highest priority (and lowest number
of flashes) is indicated. All faults latch disabling the controller and require the controller
to be turned off then on again if the source of the fault is removed.
3.1 Security Function:
1. The forward/backward/turn speed, acceleration and deceleration can be programmable.
2. Thermal overload protection.
3. Current overload protection.
4. The fault diagnostics of the controller. (Ref. Last page).
5. The power-chair will not turn on while charging the batteries.
4. Operation Guide:
1.
Rotate the clutch lever of gearbox inward to connect the drive. (paralleled to the
wheels)
2.
After seated on the wheelchair, turn on the power (Press the power button), and the
battery gauge will turn on.
3.
Start to steer, moving the joystick to drive forward, backward, left and right. If you
want to stop the wheelchair, just release the joystick, it will return to the center
position to stop the power-chair automatically.
Flash code sequence
‧-‧ -‧-
‧‧-‧‧-‧‧-
‧‧‧-‧‧‧-
‧‧‧‧-‧‧‧‧-
‧‧‧‧‧‧-‧‧‧‧‧‧-