8. For purposes of doing an initial operation check on the unit,
configure only the 5D modules that are installed for Channel 1 thru
Channel 4. Press Limits > Colors to enter the corresponding limits
and display colors as desired. Press the Analog input channel box or
the Next button to select the desired Channel. Press OK > Exit to
return to the Home page.
9. Upon return to the Home page, press Calibration > Set up a 5D
module. Select the appropriate module icon box that user would like
to calibrate. It is suggested to have a live input from the sensor that
the user can use to stimulate input signals to the 5D conditioner.
10. Within the module setup, check the input Tag name and Legend
information and enter the 5D parameters that are appropriate for
user application. Press Next and enter the parameters listed –
Transducer Sensitivity, Expected Full Scale Input and Offset. These
values are considered "absolute" calibration parameters which are
14 |
Daytronic Model 4500 Quick Reference Guide