Dynalabs DYN-PM-440 Manuale del prodotto - Pagina 13
Sfoglia online o scarica il pdf Manuale del prodotto per Apparecchiature industriali Dynalabs DYN-PM-440. Dynalabs DYN-PM-440 14. 440 n pm shaker
All rights reserved. Reproduction or issue to third parties in any form whatsoever is not permitted without written authority from the proprietors.
4) Maintenance and Troubleshooting
The DYN-PM-440 shaker is a sealed device requiring no maintenance if the
operating instructions described in this manual are followed. Repair of
damage to the coil, exciter body or magnet core should not be attempted.
The only user replaceable part is the fuse placed inside the shaker. In order
to replace the fuse unplug all cables from the shaker and remove the O-ring
as shown below. Then turn the cover in counter clockwise direction as
shown. Then replace the fuse and close the cover. For any other problem,
please return the shaker to Dynalabs for proper repair.
5) Safety
Please ensure that this manual section is reviewed and understood prior to
installation, operation or maintenance of the equipment. The danger of
electrical shock or fire always exists in electrical equipment.
The DYN-PM-440 Shaker is designed for safe operation. Safety features
such as electrical insulation on outer surfaces are provided for the safe
operation of the shaker system.