GAIMC pH100 Manuale del prodotto - Pagina 3

Sfoglia online o scarica il pdf Manuale del prodotto per Apparecchiature di prova GAIMC pH100. GAIMC pH100 3. Ph tester

pH100
Product Manual
①Remove the protective cap and then Turn the meter on by pressing the "ON/OFF" Button.
②Immerse the ph meter electrode in the solution to be tested (can not be over the immersion line) and then stir gently to wait about
30 seconds till the reading stabilized.
③After finished,clear the electrode with pure or distilled water,dry it with tissue,Turn the meter off by pressing the "ON/OFF" Button.
Note:
Advice to calibrate the meter before using, Please check the calibration chart.
CALIBRATION
pH100 has a three-point calibration function: 4.00pH, 6.86pH, 9.18pH.
6.86pH: Pour the whole package of 6.86pH powder into a dry beaker, pour into 250mL of distilled water to completely dissolve it,
and make a pH 6.86 standard solution. Press "ON/OFF" to turn on, and put the pH pen into the pH 6.86 standard solution. Long press
the "CAL" calibration key, the screen flashes 3.86 3 times, release the "CAL" calibration key, enter the automatic calibration mode,
when the display number stabilizes at 6.86pH, calibration is completed.
② 4.00pH: Pour the whole package of 4.00pH powder into a dry beaker, pour into 250mL of distilled water to completely dissolve it,
and make a pH4.00 standard solution. Press "ON/OFF" to turn on, and put the pH pen into the pH4.00 standard solution. Long press
the "CAL" calibration key, the screen flashes 6.86, quickly press the "CAL" calibration key once, the screen flashes 3.00 times,
release the "CAL" calibration key, enter the automatic calibration mode, when the display number is stable at 4.00pH, The calibration
is complete.
③ 9.18pH: Pour the whole package of 9.18pH powder into a dry beaker, pour into 250mL of distilled water to completely dissolve it,
and make a pH9.18 standard solution. Press "ON/OFF" to turn on, and put the pH pen into the pH9.18 standard solution. Long press
the "CAL" calibration key, the screen flashes 6.86, quickly press the "CAL" calibration key twice, the screen flashes three times 9.18,
release the "CAL" calibration key, enter the automatic calibration mode, when the display number is stable at 9.18pH ,
The calibration is complete.
Note: During the calibration process, the pH pen must not leave the calibration solution. If Err (as shown in the figure below) appears,
recalibration is required.
Xi 'an Gavin Electronic Technology Co., Ltd
T : +86 029-81292510 E : [email protected]
W : www
gaimc.com
.
2