Gako PM140 Manuale d'uso - Pagina 15
Sfoglia online o scarica il pdf Manuale d'uso per Apparecchiature di laboratorio Gako PM140. Gako PM140 16.
Is the air entrapped in the gel removed?
Prepare a PM jar adding a cream base and a liquid API, then close the lid of the jar
to test mixing (both gako PM jar 100ml HV and gako PM jar 125ml HV+LV
can be used).
Place the PM jar into the PM jar holder and insert it by following the steps above.
Is the PM jar fully inserted into the well?
Close the lid of the device, set the time to 45 seconds from the button and press
"START".
Does the device start operation?
Is the operation time the same as the set time?
Does the device stop operating automatically?
Open the lid of gako PM140 and remove the PM jar with the PM jar holder.
Take the PM jar out from the PM jar holder and check the preparation.
Does it look homogeneous?
Prepare a gako PM jar 125ml HV+LV adding a suppository base and an API, then close the lid of
the PM jar to test mixing.
Place the PM jar into the PM jar holder and insert it by following the steps above.
Set the time to 15 minutes from the button and press "START".
Does the device start operation?
Is the operation time the same as the set time?
Does the device stop operating automatically?
Open the lid of gako PM140 and remove the gako PM jar 125ml HV+LV
with the PM jar holder.
Take the PM jar out from the PM jar holder and check the preparation.
Does the suppository base melt?
Does the preparation look homogeneous?
Date
Date
Assessment
Signature
Signature
15
Conclusion
OK
Not OK