CLEANING/DISINFECTING AND FILTER REPLACEMENT INSTRUCTIONS (CONT.)
Step 10:
Remove a Metrex CaviWipe from its package. Begin by cleaning the surface "C". Wipe the CaviWipe
over the surface area, thoroughly wetting the soil. Then using hand pressure rub the wipe over the
surface until soil is no longer visible on this part of the device. If soil is heavy or is still visible, repeat
step with a new CaviWipe until soil is no longer visible.
Step 11:
Remove a Metrex CaviWipe from its package. Begin by cleaning the surface "E". Wipe the CaviWipe
over the surface area, thoroughly wetting the soil. Wait 1 minute, then using hand pressure rub the
wipe over the surface until soil is no longer visible on this part of the device. If soil is heavy or is still
visible, repeat step with a new CaviWipe until soil is no longer visible.
Step 12:
Remove a Metrex CaviWipe from its package. Begin by cleaning the surface "F". Wipe the CaviWipe
over the surface area, thoroughly wetting the soil. Wait 1 minute, then using hand pressure rub the
wipe over the surface until soil is no longer visible on this part of the device. If soil is heavy or is still
visible, repeat step with a new CaviWipe until soil is no longer visible.
Step 13:
Remove a Metrex CaviWipe from its package. Begin by cleaning the surface "G". Wipe the
CaviWipe over the surface area, thoroughly wetting the soil. Wait 1 minute, then using
hand pressure rub the wipe over the surface until soil is no longer visible on this part of
the device. If soil is heavy or is still visible, repeat step with a new CaviWipe until soil is no
longer visible.
Step 14:
Remove a Metrex CaviWipe from its package. Begin by cleaning the surface "H". Wipe the
CaviWipe over the surface area, thoroughly wetting the soil. Wait 1 minute, then using
hand pressure rub the wipe over the surface until soil is no longer visible on this part of
the device. If soil is heavy or is still visible, repeat step with a new CaviWipe until soil is no
longer visible.
Step 15:
Allow the device to dry, no visible wetting remains. No rinsing is required. With the
device cleaned, it is ready for disinfecting.
DISINFECTING
Disinfectants are typically provided as a concentrate to be mixed with water or ready to use (RTU) with no mixing required. Use Clorox
Healthcare Fuzion Cleaner Disinfectant RTU in a spray bottle. As an RTU product, no mixing is necessary. Clorox recommended disinfectant
wait time is 1 minute with no rinsing required.
In Steps 16 - 24, spray Clorox Healthcare Fuzion Cleaner Disinfectant liberally on each surface until that surface is thoroughly wetted.
Step 16:
Take the 8-quadrant wiping cloth from the filter carton, folding it as before so quadrant 1 is showing.
Step 17:
Spray disinfecting solution over surface A. Wait 4 minutes, ± 5 seconds, then thoroughly wipe with the cloth folded to
quadrant 1. Continue wiping until all solution has been removed.
Step 18:
Spray disinfecting solution over surface B. Wait 4 minutes, ± 5 seconds, then thoroughly wipe with the cloth folded to
quadrant 2. Continue wiping until all solution has been removed.
Step 19:
Spray disinfecting solution over surface C. Wait 4 minutes, ± 5 seconds, then thoroughly wipe with the cloth folded to
quadrant 3. Continue wiping until all solution has been removed.
Step 20:
Spray disinfecting solution over surface D. Wait 4 minutes, ± 5 seconds, then thoroughly wipe with the cloth folded to
quadrant 4. Continue wiping until all solution has been removed
Step 21:
Spray disinfecting solution over surface E. Wait 4 minutes, ± 5 seconds, then thoroughly wipe with the cloth folded to
quadrant 5. Continue wiping until all solution has been removed.
10
C
E
G
H
F