Activox LifeChoice Manuale dell'operatore - Pagina 9
Sfoglia online o scarica il pdf Manuale dell'operatore per Apparecchiature mediche Activox LifeChoice. Activox LifeChoice 16. Lifechoice activox portable oxygen concentrator
ALARMS/LIGHT INDICATORS
NORMAL OPERATION INDICATORS
Normal
Audible
Operation
Alert
LifeChoice
Brief
Activox POC
Audible
turned on
Tone
Green LEDs of Power, Active-High and
Pulse Setting LED corresponding to the
LifeChoice
None
Activox POC
has detected
an inhalation/
breath and is
delivering a
pulse of oxygen
LifeChoice
None
Activox POC
powered by
internal battery
LifeChoice
None
Activox POC
powered by
external battery
14
Light Indication
begin to use
the LifeChoice
Activox POC
Pulse Setting of 1 are illuminated.
Note: After the initial power on, the
last flow rate used will illuminate
Activox POC
(( ))
is functioning
Green LED of selected Pulse Setting
flashes with each breath
LifeChoice
Activox POC
is functioning
(( ))
Green LED light labeled INT
slowly flashes
LifeChoice
Activox POC
is functioning
(( ))
Green LED light labeled EXT
slowly flashes
Action
No breath
You may
detected by
the unit for
approximately
75 seconds
unit in
30 seconds
LifeChoice
normally
Low Battery
(Approximately
10 minutes of
battery power)
normally
Low Battery
(Approximately
30 seconds of
battery power)
normally
General
malfunction
LifeChoice
Activox POC
ALARM TABLE
Problem
Audible
Light Indication
Alert
Intermittent
beep
Amber Alarm LED light flashes
Single loud,
Alarm
long beep
every 60
seconds
Amber Battery Check LED labeled
INT slowly flashes
Critical
Fast
audible
Alarm
beep
(Unit shuts
down)
Amber Battery Check LED labeled
INT begins flashing quickly
Continuous
audible
of the
alarm
Alarm LED is solid Amber
Action
Check the
cannula
connection.
Ensure that
(( ))
you are
breathing
through your
nose. After the
system detects
a breath, the
alarm will
shut off.
If the alarm
persists,
contact your
equipment
provider.
Connect the
LifeChoice
Activox POC
to an external
(( ))
power source,
or change to
another source
of oxygen.
Connect the
LifeChoice
Activox POC
to an external
(((( ))))
power source,
or change to
another source
of oxygen.
Turn off
the unit.
Change to
another
source of
oxygen and
contact your
equipment
provider.
15