SPECIFICATIONS
ontinued
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
®
The LifeChoice
Activox Portable Oxygen Concentrator Model XYC100B is
intended for use in the electromagnetic environment specified below. The
customer or the user of the Model XYC100B should assure that it is used in
such an environment.
Immunity Test
IEC 60601
Test Level
Conducted RF
3 Vrms
IEC 61000-4-6
150kHz-80MHz
Radiated RF
3 V/m
IEC 61000-4-3
80MHz- 2.5GHz
Where 'P' is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and 'd' is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic
a
site survey,
should be less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following
symbol:
NOTE 1: At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propa-
gation is affected by absorption and reflection from structures, objects, and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and
TV broadcast cannot be predicted theoretically with accuracy. To assess the electro-
magnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the
Model XYC100B is to used exceeds the applicable RF compliance level above,
the Model XYC100B should be observed to verify normal operation. If the abnormal
performance is observed, additional measures may be necessary, such as
re-orienting or relocating the Model XYC100B.
20
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Compliance
Electromagnetic
Level
Environment - Guidance
Portable and Mobile RF
communications equipment
should be used no closer
to any part of the Model
XYC100B, including cables,
than the recommended
separation distance
calculated from the equation
applicable to the frequency
of the transmitter.
Recommended
Separation Distance
3 V
d = 1.17 √P
3 V/m
d = 1.17 √P, 80MHz
to 800MHz
d = 2.33 √P, 800MHz to
2.5GHz
b
Recommended Separation Distances Between Portable
and Mobile RF Communications Equipment and The
LifeChoice
®
Activox Portable Oxygen Concentrator Model XYC100B
®
The LifeChoice
Activox Portable Oxygen Concentrator Model XYC100B is
intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the Model XYC100B
can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters)
and the Model XYC100B as recommended below, according to the maximum
output power of the communications equipment.
Separation Distance According to the Frequency of Transmitter (m)
Rated Maximum
150kHz to 80MHz
Output Power of
d=1.17 √P
Transmitter (W)
0.01
0.12
0.1
0.37
1
1.2
10
3.7
100
12
For transmitters rated at a maximum output power not listed above, the recommended
separation distance 'd' in meters can be estimated using the equation applicable
to the frequency of the transmitter, where 'P' is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80MHz and 800MHz, the separation distance for the higher frequency range
applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects, and people.
Guidance and Manufacturer's Declaration - Electromagnetic Emissions
®
The LifeChoice
Activox Portable Oxygen Concentrator Model XYC100B is
intended for use in the electromagnetic environment specified below. The
customer or the user of the Model XYC100B should assure that it is used in
such an environment.
Emissions Test
Compliance
RF Emissions, CISPR 11
Group 1
RF Emissions, CISPR 11
Class B
Harmonic Emissions,
Class A
IEC 61000-3-2
Voltage Fluctuations/Flicker
Complies
Emissions, IEC 61000-3-3
SPECIFICATIONS
ontinued
80MHz to 800MHz
800MHz to 2.5GHz
d=1.17 √P
d=2.33 √P
0.12
0.23
0.37
0.74
1.2
2.3
3.7
7.4
12
23
Electromagnetic Environment - Guidance
The Model XYC100B uses RF energy only for
its internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
The Model XYC100B is suitable for use in all
establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.