PUMP
4
3
2
6
10441 & 10448 MATTRESS
10449 MATTRESS
(1) Pillow
(5) Male quick connector
(2) Cover
(6) Protective cap of connector
(3) Fixing belt
(7) Air tubing
(4) Low air loss holes
(8) CPR valve
(9) Foam Base
1 - Power switch
2 - Presssure adjust knob
3 - Normal pressure indicator (green)
5
4 - Low pressure indicator (amber)
5 - Static control
1
6 - Air pressure valve
1 - Pillow function cells
2 - Cover
3 - Fixing belt
4 - Low air loss holes
5 - Male quick connector
6 - Air tubing
7 - CPR Valve
1
3
8
4
7
5
6
4
STORAGE
1. Switch off power from main and unplug,
separate the pump from the mattress and
place the mattress flat.
2. Fold the mattress in two from mattress foot
to head.
3. Allow time for air to escape from mattress to
reduce storage space.
4. Close the protective cap cover on the quick
connector.
GENERAL CONDITIONS OF DISPOSAL
Dispose of the device in the appropriate
disposal areas for recycling. At the end of its
life, the product must not be disposed of along
with other domestic waste.
TROUBLESHOOTING
If your questions can't be answered with below information, please contact your
local agent directly. They might require a technician to take care of the problem.
No lights on unit
Low pressure alert
2
Patient's body is sagging
in the middle/bottoming out
Control lock up or "freeze"
9
MAINTENANCE
In case of repair, use only original spare parts
and accessories. For spare parts and
accessories, please contact Dynarex or an
authorized dealer.
CLEANING AND DISINFECTION
WARNING! Unplug the pump before
starting the cleaning procedure.
1. Clean the mattress with mild soap.
2. Do not immerse or soak pump unit.
3. After cleaning, dry the mattress without
direct sunlight exposure.
4. Check that the connection tubes are not
obstructed by dust or other objects. Use a
soft swab to clean the tubes if necessary.
NOTE: Never use acids, alkalis or solvents
such as acetone to clean the mattress.
Check control unit is connect to the main power supply and that
the unit is switched on.
Check quick blow fuse on the control unit or main plug fuse.
Check main power to the control unit and that the plug is switched.
Check mattress air supply connector is connected to the pump unit
correctly.
Check CPR connector is connected properly.
Under the mattress cover check all "T" & "L" connectors are
connected to the air tubes.
Check the connector tubes for kinks, crimps or damage.
Increase the value of the comfort the setting. Wait a couple of
minutes so pressure can stabilize before making another change.
Turn the control unit off and unplug the pump. Wait a few minutes
and plug the control unit back in. Turn on the control unit.
5