PREPARING THE NEBULIZER FOR USE
CAUTION:
To prevent the risk of cross infection, different persons should
never use the same nebulizer kit.
1. At normal use, place the device on a stable, sturdy and flat
surface horizontally, such that the unit can be easily
reached when you are seated.
Caution: place the device at least 10 cm (4 inches)
distance from wall.
2. Make sure that the unit is in the "off" (0) position
by pressing on the right side of switch.
3. Insert the power plug into the electrical outlet.
4. Rotate the inhalation top counterclockwise to remove
the inhalation top from the medication tank.
5. Add the correct amount of prescribed medication to the
medication tank.
6. Turn the inhalation top clockwise until securely closed.
7. Attach the desired inhalation accessory.
ATTACHING THE INHALATION TUBE
1. Push the Air Plug on one end of the Air Tube onto the
Air Connector on the front side of the compressor.
2. Push the Air Plug on the other end of the Air Tube onto the
Air Tube Connector on the bottom of the Nebulizer Kit.
USING THE DEVICE
1. Switch on the power.
2. Plug the inhalation tube into the
air plug under cover of device.
3. Plug the other side of tube into the
vent port of bottom of nebulizer kit.
4. Plug the nebulizer kit into the
holder on the device.
5. Add the medicine into the
nebulization tank.
6. Press the I/O button to turn on device
and begin treatment.
7. Press the I/O button to safely
terminate the operation of nebulizer.
8. Unplug the power cord from the electrical outlet
after using the device.
CAUTION
The patient and operator must keep a distance from the device
of 20 cm to 50 cm (8-20 inches) in normal use.
USING THE MASK
Place the mask over the nose and mouth. Pull the elastic strap
over the head. Gently pull on the strap to securely hold the mask
over the nose and mouth. Inhale the medication. Exhale normally
through the mask.
CLEANING
CLEANING
Following the cleaning instructions after each use will prevent any
remaining medication in the bottle from drying, which can result in
the device not nebulizing effectively.
Rinse the nebulizer parts after each use.
1. Remove the inhalation accessory (mask, mouthpiece)
from the nebulizer kit.
2. Disconnect the inhalation tubing from the nebulizer.
3. Gently twist the inhalation top counterclockwise and lift
to separate the nebulizer into two sections.
4. Remove the baffle.
5. Discard remaining medication.
6. Rinse all the applied parts (nebulizer kit, the mouthpiece and
mask) in warm water.
7. Air dry only.
8. CLEANING THE DEVICE AND THE TUBE'S OUTER SURFACE
Only use a damp cloth.
9. Assemble the nebulizer and store the nebulizer kit in a dry,
sealed bag.
CAUTION: The nebulizer kit should be replaced every 6 months.
4
CARING FOR THE DEVICE
To keep your device in the best condition and protect the unit
from damage, follow these directions:
CLEANING THE COMPRESSOR
Clean the casing of the main unit by using a soft cloth
moistened with water or a mild detergent. Do not use abrasive
cleaners. Dry the casing immediately using a soft clean cloth.
CHANGE THE AIR FILTER
Change the air filter every 60 days even if the air filter does
not appear dirty. If the air filter appears dirty, or if water or
medication has spilled on the air filter, replace with a new air
filter immediately.
1. Pull the air filter cover to remove from the front side
of the compressor.
2. Remove the dirty filter by hand.
CAUTION: Do not attempt to wash or clean the air filter. Damp
air filters can cause blockages. Do not substitute cotton or any
other material for the air filter.
CAUTION: Wash the air filter cover regularly to prevent any
blockage in the cover. Do not boil. Make sure the cover is dry
before inserting the new air filter.
3. Insert a new air filter into the air filter cover.
CAUTION: Before inserting the new air filter, make sure the air
filter is clean and free of dust. Do not operate the device without
the air filter.
4. Put the air filter cover back on the compressor.
5. Do not maintain or service the device while it is in use.
TROUBLESHOOTING GUIDE
PROBLEM
CAUSE
No power on unit
The AC power cord is not
Turn the power switch off. Plug
when the power
plugged into an electrical
the power plug into an electrical
switch is on.
outlet.
outlet. Turn the device on.
No medication in the medi-
Add the correct amount of
cation tank. Too much or
prescribed medication to the
too little medication in the
medication tank.
medication tank.
The nebulizer kit is not
Make sure the nebulizer kit is
correctly assembled.
correctly assembled and the
inhalation accessory is correctly
attached.
No nebulization or
The nebulizer kit is tilted at
Hold the nebulizer kit correctly.
low nebulization
an incorrect angle.
Do not tilt the nebulizer kit so the
rate when the
angle of the kit is greater than
power is on.
45 degrees.
The air tube is incorrectly
Make sure the air tube is correctly
attached to the compressor and
attached.
the nebulizer kit.
The air tube is folded or
Make sure the air tube is not
damaged. The air tube is
folded, kinked or bent. Inspect the
blocked.
air tube for any damage. Replace
the air tube if damaged.
The device is
The compressor is
Do not cover the compressor with
very hot.
covered. The device has
any type of cover during use. Turn
been used for longer than
the device off. Wait 50 minutes
50 minutes.
before using the device again.
WARRANTY
Your Dynarex Resp-O
Elite Compressor Nebulizer is warrantied to
2
be free of defects in materials and workmanship for three (3) years
from original date of purchase, for use by the original purchaser.
This item was built to exacting standards and carefully inspected
prior to shipment. This Limited Lifetime Warranty is an expression
of our confidence in the materials and workmanship of our products
and our assurance to the consumer of dependable service.
The Warranty shall not apply under the following conditions:
• Problems arising from normal wear
• Problems arising from failure to adhere to the product
instructions
• Problems arising from misuse, negligence, accident or
improper operation, maintenance or storage
• Problems arising from modifications or unauthorized repairs,
parts or attachments
• Products where the serial number has been removed or
defaced
• Problems with non-durable components, such as rubber
accessories, casters, and grips, which are subject to normal
wear and need periodic replacement
Dynarex shall not be liable for any consequential or incidental
damages whatsoever. Dynarex shall repair or replace defective
products at its option. The foregoing warranty is exclusive and
in lieu of other express warranties, if any, including the implied
warranties of merchantability and fitness of a particular purpose.
The remedy for any violation of the implied warranty shall be
limited to repair or replacement of the defective product pursuant
to the terms contained herein. If you have a question about your
Dynarex device or this warranty, please contact an authorized
Dynarex dealer.
5
Equipment Classification with respect to protection
from electric shock: CLASS II
Degree of protection from electric shock: TYPE BF
Degree of protection against ingress of water is rated as IPX0.
Equipment not suitable for use in the presence of flammable
anesthetic mixture with air, oxygen or nitrous oxide.
Intermittent operation: 50 minutes ON, 50 minutes OFF.
Input
SOLUTION
Extreme Pressure
Power Consumption
Medication Capacity
Particle Size
MMAD
Average Nebulization Rate
Noise Level
Compressor Pressure Range
Operation Pressure Range
Liter Flow Range
Operation mode
Operation Temperature Range
Operation Humidity Range
Storage Temperature Range
Storage Humidity Range
Accessories
Dimensions
NOTES:
• Subject to technical modification without prior notice.
• Please note that specifications may vary with medication
type used.
• Do not use the device where it may be exposed to flammable gas.
• This unit conforms to EMC Standard IEC60601-1. However, if it is
used together with other medical devices or electrical equipment,
they may influence the operations of one of devices. Please follow
any instructions in the manuals and use all devices correctly.
Dynarex Resp-O
shock, fire and mechanical hazards only in accordance with
IEC60601-1
Manufactured for:
Dynarex Corporation
10 Glenshaw Street
Orangeburg, NY 10962
USA • www.dynarex.com
Made in China R210429
149°F
65°C
14°F
-10°C
10%
CLASSIFICATION
SPECIFICATIONS
AC 120 V, 60 HZ
>205 kpa
≤200 VA
10 mL
≤5 µm & 60%
2.0 µm
≥0.2 mL/min
Below 60 dB
30-58 psi
14.5-27 psi (1 bar-1.9 bar)
16 ~ 23.5 lpm
50 minutes on, 50 minutes off
10˚C to 40˚C (50˚F to 104˚F)
10 to 95% RH
-10˚C to 65˚C (14˚F to 149˚F)
10 to 95% RH
1 Adult Mask, 1 Pediatric Mask,
1 Mouthpiece, 1 T-Piece,
1 7' (2.1m) Tube, 5 Filters
290 mm × 180 mm × 110 mm
Elite Compressor Nebulizer with respect to electric
2
SYMBOL GLOSSARY
For an explanation of symbols
used in Dynarex packaging, visit
dynarex.com/symbols.php
IPX0
95%
CAUTION: Federal law (USA)
restricts this device to sale by
or on the order of a physician.
6