iHealth BP7 Manuale d'uso - Pagina 8

Sfoglia online o scarica il pdf Manuale d'uso per Monitor della pressione sanguigna iHealth BP7. iHealth BP7 16. Wrist
Anche per iHealth BP7: Manuale d'uso (16 pagine), Manuale di avvio rapido (4 pagine)

iHealth BP7 Manuale d'uso
d) Simultaneous use with other medical monitoring equipment on the same limb
e) The blood circulation of the user needs to be checked
3.
This Wireless Blood Pressure Wrist Monitor is designed for adults and should never be used
on infants, young children, pregnant or pre-eclamptic patients. Consult your physician before
use on children.
4. Do not use this product in a moving vehicle as this may result in inaccurate measurements.
5. Blood pressure measurements determined by this product are equivalent to those obtained by
professional healthcare practitioners using the cuff/stethoscope auscultation method within the
limits prescribed by the American National Standard, Electronic or Automated
Sphygmomanometer.
6. Information regarding potential electromagnetic or other interference between the blood
pressure monitor and other devices together with advice regarding avoidance of such
interference, please see ELECTROMAGNETIC COMPATIBILITY INFORMATION. It is
suggested that the blood pressure monitor be kept 10 meters away from other wireless devices,
such as WLAN unit, cell phone, microwave oven, etc.
7. It is recommended that the iOS device be set in Airplane mode during measurement to avoid
strong magnetism interference. If a call comes in during the measurement, the measurement
process will be terminated automatically.
8. If Irregular Heartbeat (IHB) is detected during the measurement procedure, the IHB symbol
will be displayed. Under this condition, the Wireless Blood Pressure Wrist Monitor can keep
functioning, but the results may be inaccurate. Please consult your physician for accurate
assessment.
There are 2 conditions under which the signal of IHB will be displayed:
1) The coefficient of variation (CV) of pulse period >25%.
2) The difference of adjacent pulse period ≥0.14s and the number of such pulse takes more than
53 percent of the total number of pulses.
9. Please do not use any other cuff other than that supplied by the manufacturer as this may result
in measurement errors and a biocompatible hazard.
10.
This product might not meet its performance specifications if stored or used outside the
specified temperature and humidity ranges.
11.
Please do not share the cuff with any infectious person to avoid cross-infection.
12. This product should not be used as a USB device.
13. This product is verified by auscultatory method. It is recommended that you check Annex B of
ANSI/AAMI SP-10:2002+A1:2003+A2:2006 for verification method details if needed.
14. If the determined blood pressure (systolic or diastolic) is outside the rated range specified in
part SPECIFICATIONS, the app will immediately display a technical alarm on screen. In this
case, consult a physician or ensure that proper measurement procedures are followed. The
technical alarm is preset in the factory and cannot be adjusted or inactivated. This technical
alarm is assigned as low priority according to IEC 60601-1-8.The technical alarm is
non-latching and does not need to be reset.
15. A medical AC adapter with an output of DC 5.0V and complies with IEC 60601-1/UL 60601-1
and IEC 60601-1-2/EN 60601-1-2 is suitable for this monitor, such as ASP5-05010002JU
(input: 100-240V, 50/60Hz, 200mA; output: DC 5V, 1.0A). Please note that the monitor jack
size is USB mini B.
7