This diagram is intended to illustrate the operation of commands involved
in the calibration process.
The Calibration Sequence is:
•
Send 0CV
•
Turn the compass so it points to 0 i.e. it is pointing magnetic North
•
Rotate the A4032 until the output angle is zero
•
Send a sequence of PCV commands while the compass is rotated
– one every 10deg
•
Send SCV to update the table used for correction
3.2 Use of PC Calibration Software Tools
This describes the operation of the two Autonnic A4032 Calibration
software tools
Factory Utility 1.13
P6033 Calibration Tool V1.02
[Note that the Calibration Tool is only for use with A4032 firmware
v040 and later]
This normally is delivered packed as a zip file: A4032 Cal tool kit.zip
This must first be unpacked so it can be used. Once that is done the files
can be installed.
Open the Factory Utility 1.13 directory and use the file called setup.exe
This is likely to cause your PC to advise against it – but ignore the
warnings and complete the installation
Similarly, open the P6033 Calibration Tool V1.02 directory and use the file
also called setup.exe to install it.
This is also likely to cause your PC to advise against it – but, again, ignore
the warnings and complete the installation.
Hardware notes:
1 The A4032 is to be attached to a magnetic compass which itself can
be rotated while the A4032 does not move relative to the
compass so that, at all times, the A4032 data represents the
angle of the magnet of the Compass.
2 The PC will have a port assigned to the USB-RS232 cable attached
to the A4032 and the number of this port must be found from
the Device Manager Windows utility. Eg COM 6