DCbox GBMA Manuale - Pagina 2

Sfoglia online o scarica il pdf Manuale per Monitor DCbox GBMA. DCbox GBMA 2. 4" (10 cm) dot matrix large display

DCbox GBMA Manuale
Display
ROP
Alarm Setting
The following steps are only available for alarm output.
Page (roP)
ENT
Press
AL1
Alarm 1
Press
Setpoint (AL1)
ENT
Press
AL2
Alarm 2
Press
Setpoint (AL2)
ENT
Press
AL3
Alarm 3
Press
Setpoint (AL3)
ENT
Press
AL4
Alarm 4
Press
Setpoint (AL4)
ENT
Press
Alarm 1
AC1
Press
(ACt1)
>
(Hi) or <(Lo) for alarm action.
Alarm 2
=
AC2
(ACt2)
Press
AC3
Alarm 3
>
(Hi) or <(Lo) or (Go) for alarm action.
=
(ACt3)
AC4
Press
Alarm 4
(Hi) or <(Lo) or (Err) for alarm action.
(ACt4)
ENT
Press
Hysteresis 1
Press
HY1
(HYS1)
lower or higher display value (depends on alarm action).
Hysteresis 2
Alarm setpoint this range (0~999) will turn off the alarm.
HY2
PS: 1. There are 4 alarms output optional.
(HYS2)
2. This page is exist without alarm output, but the
Hysteresis 3
HY3
function will be disabled.
(HYS3)
3. Press ENT to save the value and go to the next
Hysteresis 4
HY4
parameter.
(HYS4)
Delay Time 1
ENT
Press
Press
DE1
reach the alarm value that need to wait for this time
(dEL1)
(0~99 sec) for alarm action.
Delay Time 2
DE2
(dEL2)
PS: 1. There are 4 alarms output optional.
2. This page is exist without alarm output, but the
Delay Time 3
DE3
function will be disabled.
(dEL3)
3. Press ENT to save the value and go to the next
DE4
Delay Time 4
parameter.
(dEL4)
ENT
Press
Alarm Start Band
Press
SB
value don't over this range; the alarm will not be act.
Setting (Sb)
ENT
Press
Press
Alarm Start Band
SDT
value reach alarm start band value; the alarm will be act
Time Set ting (Sdt )
after this value (sec).(The function is used with "Sb" function.)
ENT
Press
AOP
A/O Setting Page
The following steps are only available for analog output.
(AoP)
ENT
Press
Press
A/O Polarity
POL
PS : Voltage output ,NO: positive pole output (0~+10V)
Setting (PoLAr)
ENT
YES: positive & negative pole output (-10~+10V)
Press
Press
A/O Low Scale
ANL
the display value (programmable).
Setting (AnLo)
ENT
Press
EX : A/O is 0~10V, the display is 10.0 to output 0V, this value must be set for 10.0.
ANH
Press
A/O Hi Scale
the display value (programmable).
Setting (AnHi)
ENT
Press
EX : A/O is 0~10V, the display is 90.0 to output1 0V, this value must be set for 90.0.
Descriptions
Alarm Setting Group Procedures
to modify alarm 1 setpoint.
to modify alarm 2 setpoint.
to modify alarm 3 setpoint.
to modify alarm 4 setpoint.
to modify alarm value that is
to modify alarm value that is
to modify alarm value that is
to modify the value, when alarm runs
+ -
to modify the value, when the display value
to modify the value (-99~+99), if the display
to modify the value (0~99 sec), if the display
A/O Setting Group Procedures
to select output for positive or negative pole.
to adjust A/O low scale to correspond to
to adjust A/O hi scale to correspond to
P3
Default
Display
DOP
RS485 Setting
Page (doP)
ENT
Press
0000
ADD
Address Setting
(Addr)
ENT
Press
0000
Baud Rate
BAU
Setting (bAUd)
ENT
Press
0000
Parity Setting
PAR
(PAri)
0000
ENT
Press
Frame Setting
FRA
(FrAmE)
HI
ENT
Press
ADJ
HI
ENT
Press
DOF
Display Offset
HI
Setting (doF)
ENT
Press
Display Gain
DGA
Setting (dGA)
ENT
Press
0000
A/O Offset
AOF
Setting (AoF)
ENT
Press
AGA
A/O Gain
Setting (AGA)
ENT
Press
00
Display
IO
-IO
Input signal is under -20% of input range.
00
DO
Input signal is over 180% of input range or meter error.
-DO
00
ADE
Input signal is under display range (-19999)
E00
EEPROM reading/writing suffers the interference (about 1 million times).
**Please check the wiring connection is correct first, if the problem still exist, please return the meter
to the factory.
NO
0000
9999
Descriptions
RS485 Setting Group Procedures
The following steps are only available for RS-485.
Press
to modify address (0~255).
Press
to select baud rate (38400/19200/9600/4800).
Press
to select parity (n.8.2/n.8.1/even/odd).
Press
to select frame type.
( NO:Hi
Lo , YES:Lo
Hi )
Input / Output Adjustment Procedures
Press
to display offset value (-1999~9999).
Press
to display gain value (-1999~9999).
Press
to analog output offset value (-1999~9999).
Press
to analog output gain value (-1999~9999).
Error Code of Self-Diagnosis
Descriptions
Input signal is over 120% of input range.
Input signal is over display range (99999)
P4
Default
0000
384
n.8.2.
NO
0000
0000
0000
0000