GaleMed AR0056 Manuale di istruzioni - Pagina 5

Sfoglia online o scarica il pdf Manuale di istruzioni per Prodotto respiratorio GaleMed AR0056. GaleMed AR0056 11.

Operating Instruction 
 
 
 
 
 
 
 
 
 
 
 
 
 
5.
Hold mask firmly in place over 
the mouth and nose with 
thumb and index finger, using 
other fingers to grip the chin. 
 
6.
Inflate the lung by 
compressing resuscitation bag 
with the other hand. Develop 
rhythmic pattern for 
adequate inspiratory/ 
expiratory time. 
7.
Check to ensure that you are 
performing ventilation 
properly 
 
Observe rise and fall of the 
patient's chest. 
Check the patient's lip and 
face color through the 
transparent face mask. 
Check that the patient valve 
is working properly through 
the transparent housing. 
Check that the interior of 
the mask is being "fogged" 
General Instructions for Operation 
Select the appropriate size of MR‐100
Child or Adult). Select the appropriate size of face mask If ventilation by 
mask is to be performed. Connect mask to the outlet on non‐rebreathing 
valve.   
 
If supplemental oxygen is to be used, connect supply tubing (not supplied) 
between the regulated gas source and the oxygen reservoir inlet. Adjust gas 
flow so that the reservoir expands completely during the delivery of the 
breath and nearly collapses as the bag refills. 
If supplemental oxygen is not to be used, remove the reservoir. Failure to do 
so will affect the refill rate and maximum frequency capabilities.   
 
Should the non‐rebreathing valve become contaminated with vomit, blood 
or secretions during ventilation, disconnect the Resuscitator from the 
patient and clear the non‐rebreathing valve as follows: 
Squeeze the silicone bag to deliver several sharp breaths through the non‐
rebreathing valve to expel the contamination. If the contamination does not 
clear, continue ventilation with another resuscitator or use mouth to 
mouth/mask techniques.    Disassemble the non‐rebreathing valve and rinse 
it with water. Reassemble and disinfect. 
 
Once the procedure is completed, clean, disinfect and test the Resuscitator 
as described under Cleaning & Disinfection (P.14) & Testing the Resuscitator 
(P.15). 
 
™ 
Plus    Manual Resuscitator (Infant,