GaleMed AR0097 Manuale di istruzioni per l'uso - Pagina 7
Sfoglia online o scarica il pdf Manuale di istruzioni per l'uso per Apparecchiature mediche GaleMed AR0097. GaleMed AR0097 12. Resuscitator silicone reservoir bag, 2l
and/or patient.
3. Squeeze the resuscitator bag several times to assure that the unit is functioning properly.
4. Select the Peep Valve with appropriate range (2 ‐10cm H
5. Set the Peep Valve knob to the approximate value shown on the Peep Valve housing.
6. Connect the Peep Valve to the Flow Diverter port as shown in the illustration. Connect
the resuscitator to a manometer and test lung. Ventilate the test lung and adjust the Peep
Valve as required to obtain the desired end expiratory pressure.
7. Ventilate the patient following ACLS procedures for ventilation. Watch the chest rise
during ventilation. Also, during ventilation, periodically check for:
‐ signs of cyanosis;
‐ adequacy of ventilation;
‐ proper airway pressure;
‐ secure connection of Peep valve and diverter.
8. Between patient use or periodically, clean and sterilize the Peep Valve and Flow Diverter.
‐ Testing the Resuscitator
3.2 System Overview
3.2.1 Intended for use together with other devices:
Oxygen Tubing, Airway, Silicone Mask, Aircushion Mask, PEEP Valve, Disposable manometer
3.3 Specification/Functions
3.3.1 Operating Environmental Temperature Limits:
‐18°C to +50°C
3.3.2 Storage Environmental Temperature Limits:
‐40°C to 60°C (‐104°F to 140°F)
3.3.3 Range of setting:
Valve resistance
Oxygen
concentration
Bag capacity (ml)
Max delivered
volume (ml)
Safety pressure‐
relief valve
Reservoir bag (ml)
Mask size
Durable PEEP Valve
REF
SAP REF
2403
AV0035
Adult Resuscitator
1500
1350
60±10 cmH
O
2
2000
#4
Product name
Durable PEEP Valve
O or 5‐20cm H
2
Child Resuscitator Infant Resuscitator
<4.0 cmH
O
2
>95% @15 LPM
550
350
40±5 cmH
O
2
1000
#3
Pressure range Connector size
2‐10cmH2O
30F (Orange)
O)
2
280
160
40±5 cmH
O
2
500
#0
Package
1 pc/pkg, 50
pcs/box