SERIES 5 SIDE DELIVERY SYSTEM
Instructions For Use

ADJUSTMENTS

UTILITY CENTER
The Utility Center comes factory preset at 40 psi for Water pressure and 80 psi for Air pressure. All regulator adjustments
should be made with the Master On/Off in the ON position.
1. TO INCREASE AIR AND/OR WATER PRESSURE: Turn the
knob clockwise to the desired pressure.
2. TO DECREASE AIR AND/OR WATER PRESSURE: Turn the
knob counterclockwise to the desired pressure.
NOTE: When decreasing pressure, pressure must be relieved from the system. This may be achieved by pressing the
syringe buttons to attain an accurate measurement from the gauges.
DRIVE AIR
1.
Identify the adjustment knobs for
controlling drive air on the underside of the
control head as shown (right).
2.
Install a bur in the handpiece that is to be
adjusted. The drive air adjustment screws
on the handpiece block correspond with
the handpiece positions on the holder bar.
3.
Run the handpiece. With the foot
control plate fully depressed, turn
the corresponding adjustment screw.
Clockwise to decrease pressure,
counterclockwise to increase pressure.
10
Pressure Adjustment Knob
DRIVE AIR FLOW ADJUSTMENT
FLOW CONTROL
Pressure Gauge
DRIVE AIR
DRIVE AIR
PRESSURE GAUGE