Quick Start Guide
2
Install
2.1
Install the sensor
Procedure
1. Mount the sensor in a vertical pipe run with flow from bottom to top.
If the sensor must be mounted in a horizontal pipe run, orient the
sensor in the 3 o'clock or 9 o'clock position.
2. Orient the sensor within 80 ° of vertical. See
Figure 2-1: Sensor Orientation
8
June 2019
Figure
2-1.
Emerson.com/Rosemount