Decagon Devices MAS-1 Manuale d'uso - Pagina 17

Sfoglia online o scarica il pdf Manuale d'uso per Accessories Decagon Devices MAS-1. Decagon Devices MAS-1 20. 4-20 ma soil moisture sensor
Anche per Decagon Devices MAS-1: Manuale dell'operatore (18 pagine)

Decagon Devices MAS-1 Manuale d'uso

6. Troubleshooting

If you encounter problems with the MAS-1, they will usu-
ally be caused by one of two situations
If the MAS-1 readings in air and/or water are outside the
ranges given in the Testing the Sensor section, then
there is likely a problem with the connection to the PLC
or other data acquisition system. Check the wiring and
check to make sure that the supply voltage is in the
specified range.
If the MAS-1 is reading a negative value for VWC
while it is inserted into the soil, make sure that you have
good sensor-to-soil contact. When inserted, the MAS-1
should be completely covered up past the black over-
molding. Removing and re-installing the full length of
the sensor with good sensor-to-soil contact should rem-
edy this problem.
If problems persist, contact Decagon for assistance.
MAS-1 Soil Moisture Sensor
14
6. Troubleshooting