Deckma OMD Series Manuale di manutenzione e pulizia - Pagina 9
Sfoglia online o scarica il pdf Manuale di manutenzione e pulizia per Strumenti di misura Deckma OMD Series. Deckma OMD Series 13. Cleaning procedure of measuring cell mz-09
DECKMA
HAMBURG
1.4 Cleaning of Iron Oxide and Calcification
Step 1.
Figure 10
To clean Iron Oxide deposit, calcification,
and
other stubborn deposit, it is
recommended to use the Citric Acid from
the Service Kit (Order No.: 18570).
1. Ensure that the valves remain closed.
2. Mix some Citric Acid with 15 ml water.
3. Fill the Sample Glass Tube with the Citric
Acid liquid and let it soak over night.
4. Flush with clean water without gas.
Page: 9