Centurion Scientific C2006 Manuale di istruzioni - Pagina 5
Sfoglia online o scarica il pdf Manuale di istruzioni per Apparecchiature di laboratorio Centurion Scientific C2006. Centurion Scientific C2006 6. Led benchtop centrifuges
Above shows 12000 (Rpm see dot bottom right) IE: 1200 x 10 and 30 minutes of time
To set speed in Rpm
Press up or down arrows below speed display.
To set in Rcf (G)
Press Function button (keep held down) then press Lid open button.
NOTE the direction of blue dot in the speed display
Top left for RCF program, bottom right for Rpm.
Above shows Acceleration rate of 9, Decelleration rate of 9 and Memory number 1
To set:
acceleration or deceleration rates
Press Function button (keep held down) then
Press Speed Down arrow for acceleration
Press Speed Up arrow for Decelleration (brake)
Both show 0-9 inclusive, 9 is the highest rate
To use the centrifuge
Load samples, Fit lid (if applicable) Set Speed or Rcf(G) then designated time.
Close the Centrifuge lid. Press down firmly on both left & right sides.
Press Start button (bottom left), the motor will now accelerate to the designated setting.
When time has counted down (time display right) the rotor will decelerate (brake).
According to the setting you have made (9 is fastest) the rotor will stop and OpEn will appear on the display.
Press Lid open button and retrieve your samples.
You can stop the run at any time by pressing the Start / Stop button.
To use Pulse (short run)
This method allows very short runs mainly for "Pelleting" purposes.
Set your desired speed or RCF (G) Time is not relevant.
Load your samples. Fit lid (if applicable).
Press Pulse button (bottom right) keep held down, you will see the Time display counting up in seconds.
As long as you keep the button held down time will increase. On releasing the button centrifuge will brake.
Press Lid open when
OpEn
To recall memory
First recall memory number (see above)
Press Function button (keep held down) then
Press Time up arrow to recall memory
Follow safety instructions on previous pages. Always load samples evenly.
appears on the display , and retrieve your samples.
To set time
Press up or down arrows below time display.
To set:
memory and store
(first set speed or rcf &time)
Press Function button (keep held down) then
Press Pulse button to select number 0-9
inclusive
Press Time down arrow to store