Sfoglia online o scarica il pdf Manuale di istruzioni per Apparecchiature di laboratorio CORNING LSE 6876-SB. CORNING LSE 6876-SB 12. Digital dry bath heater
1.0 General Specifications
Temperature range
Temperature
display resolution
Temperature
uniformity
Temperature
accuracy
Temperature
controller
Timer
Operating altitude
Control
I/O
Dimensions
(W x D x H)
Weight
Electrical
requirements
2.0 Safety Precautions
Do not use this product in an explosive environment.
Do not use in the presence of flammable or combustible material.
Do not heat substances that react violently when heated.
Do not touch block when hot or when unit is heating. Use block lifter.
Do not touch area around block or block well when unit is hot.
Do not spill liquids into the well area or into the unit side vent holes.
Connect unit only to a properly grounded outlet.
2
Ambient +5°C to 150°C
0.1°C 4 digit LED
±0.2°C (at 37°C in block)
±0.3°C
Microprocessor – user calibratable
1 to 99 hours 59 minutes or continuous in
1 minute increments
Not to exceed 2,000 meters
PI Microprocessor controller
USB unidirectional
Single and dual block models:
8.3 x 12.4 x 4.7 in. (21 x29 x 12 cm)
Four block models:
8.3 x 16.3 x 4.7 in. (21 x 39 x 12 cm)
Single and dual block models:
7.0 lbs. (3.2 kg)
Four block models:
9.6 lbs. (4.4 kg)
6875-SB: 115V, 50/60Hz, 1 A, fuse 1.6 AT
6876-SB: 230V, 50/60Hz, 0.5 A, fuse 1AT
6877-SB: 230V, 50/60Hz, 0.5 A, fuse 1AT (UK plug)
6878-SB: 100V, 50/60Hz, 1 A, fuse 1.6 AT
6885-DB: 115V, 50/60Hz, 1.9A, fuse 3.15AT
6886-DB: 230V, 50/60Hz, 0.95 A, fuse 1.6AT
6887-DB: 230V, 50/60Hz, 0.95 A, fuse 1.6AT (UK plug)
6888-DB: 100V, 50/60Hz, 1.9A, fuse 3.15AT
6895-FB: 115V, 50/60Hz, 3.73 A, fuse 5 AT
6896-FB: 230V, 50/60Hz, 1.86 A, fuse 3.15 AT
6897-FB: 230V, 50/60Hz, 1.86 A, fuse 3.15 (UK plug)
6898-FB: 100V, 50/60Hz, 3.73 A, fuse 5 AT