e-survey E100 Manuale d'uso - Pagina 11
Sfoglia online o scarica il pdf Manuale d'uso per Ricevitore e-survey E100. e-survey E100 15. Gnss receiver
E100 User Manual
3.4 E-bubble Calibration
E100 is equipped with e-bubble sensor which supports tilt survey in SurPad4.0
software. To calibrate the e-bubble, put the device on flat table or pole (ensure the
bubble on the pole is normal before calibration, then centering the pole bubble). In
SurPad4.0 software, connect device and click "Device" -> "Device Settings", open "E-
Bubble" function. Then, go to "Device" -> "Calibrate Sensor", click "Calibrate" to
calibrate the e-bubble.
To use tilt survey function, go to "Survey" -> "Point Survey" page, select "Tilt Point".
Then click survey button to start data collection. After collect three points on the same
location, the software will calculate a final result.