Aliyiqi HF-2 Manuale di avvio rapido - Pagina 11

Sfoglia online o scarica il pdf Manuale di avvio rapido per Apparecchiature di prova Aliyiqi HF-2. Aliyiqi HF-2 15. Hf digital display push-pull force gauge

a. Set the machine stop value for this instrument first: Setting method: Turn on, when the measurement interface
is displayed, press \"Settings\" to enter the machine stop value setting (display
value to the preset value.
b. There is a signal output jack on the right side of this instrument (see \"Outline Structure Diagram \" above).
Insert one end of the signal output line into this jack (socket) and the other end into the electric test bench
produced by our company The jack (socket) marked as signal input on the top (this signal output line is the
standard accessory of the electric test bench).
c. Starting the electric test bench produced by our company can realize the function that the motor stops
automatically when the test force exceeds the stop value of the machine.
(LOSET) Minimum saved value of peak value:MIN peak save value. In peak mode, when the
current value is less than this value, the peak value will not be saved.
(OFFT) Automatic shutdown time setting: under this setting item, use \"
10 minutes to 90 minutes to automatically shut down, you can also set \" 00 \"does not automatically shut down.
If you do not choose to shut down automatically, press \" Save \"after selection to complete the setting. That is
Return to the option interface. The default setting of the instrument is 10 minutes.
(PE.SET) Automatic peak save time: automatic peak save time, the unit is second, the default
value is 10, the setting cannot set this value to 0.
(G.SET)Gravity acceleration setting: The user can set the gravity acceleration value according to
the location of the region, the default value is 9.800.
(BACSET) Backlight function setting: under this setting item, use \"
\"
(YES) \"means turn on the backlight function, choose \"
press \" Save \"to save and return to the setting item interface after selection.
(REF) Sensor reference value: manufacturer reference value, users do not need to pay attention.
(DIr) Positive and negative push-pull direction: used to switch the positive and negative direction of the
push-pull force gauge.
(SHIELD) Cancel the mask: Cancel the mask to display data below 1% of full scale. Press \"YES \"
to not cancel, press \"NO \" to cancel.
(RESET) Factory reset function:Under this setting item, press \"Settings \" to restore the factory
settings, and the machine will shut down. If you use the machine to restart the machine.
Nine, test
Press \"Power On \" key to turn on the power, and use the factory default settings as needed to directly test or
press the Set key to select the test mode and test.
1. Select the appropriate test joint fixture to be installed on the push-pull force gauge (for self-made fixtures,
please refer to the relevant data in \"Outline and Installation Dimensions \".
2. Please hold the push-pull force gauge firmly or install the push-pull force gauge on a suitable test machine for
testing. During the test, please make the tested force and the push-pull rod of the push-pull force gauge in line to
measure the accurate load .
After (STOP), set its
\"Key selection, you can set
\"Key selection, if selected
(NO) \"means to turn off the backlight,