Aliyiqi HF-5 Manuale di avvio rapido - Pagina 7

Sfoglia online o scarica il pdf Manuale di avvio rapido per Apparecchiature di prova Aliyiqi HF-5. Aliyiqi HF-5 15. Hf digital display push-pull force gauge

①, three kinds of display: after booting, when there is no character display, it means \"track mode \" (real-time
load measurement mode), the value on the screen changes with the change of load; press \"peak \" key here
When \"PEAK \" is displayed, it indicates Peak mode (peak hold measurement mode), and the display shows the
peak value until it is manually cleared; press \"Peak \" key here, when \"AUTO PEAK \" is displayed, it means
\"autopeak \" (Peak hold automatic release measurement mode), automatically save and clear the peak value
after displaying the peak value for 10 seconds, the peak time of 10 seconds can be freely set. For details, please
refer to the
(PE.SET) Explanation}.
② With or without data storage display: when there is no \"MEM \" displayed, it means that no data is saved in
the instrument; when there is \"MEM \" displayed, it means that there is data saved in the instrument; When you
press \"View \" on the measurement interface to view the saved data, \"MEM \" is flashing.
③ Push and pull instructions: In this machine, the thrust (pressure) is displayed as a negative value (\"-\"), and
the pulling force is a positive value (\"+ \" is not displayed).
④ Comparison function. When setting the comparison value via STOP (for details, please refer to the
(STOP) Machine stop value description), this function is activated. \"CMP \" is displayed. It can be set separately
according to need when using.
⑤, measurement force value display.
⑥ Display of upper and lower limit values reached: When \"MIN \" is displayed here, it indicates that the
measured data reaches and is below the lower limit; when \"MAX \" is displayed here, it indicates that the
measured data reaches and The upper limit is exceeded.
⑦, three units display: N (cow), kg (kilogram), lb (pound) three units are displayed separately, converted by
pressing the unit key. When HF10K and above specifications, K N (kilo-newton), K Kg (ton), \"K lb (kilo-pound) \"
are displayed here.
⑧, power display: when the battery voltage drops below 7.0V, it indicates that the voltage is insufficient and
needs to be recharged (it can still be tested during charging). There is a power indicator light above the power
button. When the charger is plugged in for charging, the indicator first changes from red to green. This process
is repeated twice, indicating that the charging is normal, then the indicator light is red, and the indicator light
changes when fully charged Into green. If the instrument is not used for a long time, and the battery power is too
low, the indicator light may not light up when the charger is plugged in. After about half an hour, after the battery
is woken up, the indicator light will turn red and the charging time will be 4 ~ 6 hours.
6. Key introduction