A-TECH MES SQA-Vt Bollettino tecnico

Sfoglia online o scarica il pdf Bollettino tecnico per Apparecchiature mediche A-TECH MES SQA-Vt. A-TECH MES SQA-Vt 3. Sperm analyzer

Problem  description: 
The  lower,  operational  screen  of  the  SQA  is  blank,  but  lit  OR  it  displays  black  lines  on  a  yellow 
background.  
Procedure  
STEP 1:  
1. Re‐install the SQA‐V software. 
2. If the software is successfully installed and the screen is still blank – go to Step 2 
3. If the software was not installed successfully based on: 
a.
ACK error:  This is a FLASH MEMORY failure. The SQA‐V must be sent back to MES LTD for repair as an RMA. 
b. TIME OUT error:  This is a communication failure between the SQA and the PC.  Verify that the SQA‐V is 
connected to the correct COM port on your PC and also that the COM port is defined properly.  
If the port is functioning: Press the SERVICE button while sequentially switching the device on and off twice. Try 
c.
to install the software again. If the installation fails again –go to Step 2.
STEP 2:   
1. Make sure that the SQA‐V is turned off and 
dissconnected from any source of electricity. 
2. Open the SQA‐V. 
3. Verify that the harness that connects the 
user screen and the main board is in the 
correct position and is properly connected 
(see Figure 1).  
4. If the problem persists go to Step 3. 
 
 
 
STEP 3:  
1. Verify that the harness connecting  the main 
board to the PC is properly connected (RS232 
cable).  
2. Tighten the harness connector at location J5 
on the main board (see Figure 2).  
3. If the problem persists go to Step 4. 
STEP 4:  
1. Verify that the Main board processor is positioned according to the instructions in Appendix 1 of this bulletin.  
2. Replace the processor if the problem persists.  
3. If the screen is still blank after replacing the processor‐ go to Step 5. 
STEP 5:  
1. Replace the LCD operational screen according to the instructions in Appendix 2 of this bulletin. The instructions can also 
be found on Page 19 in the SERVICE MANUAL.  
2. If the problem persists with the new screen – send the SQA back to the manufacturer (MES) for a repair RMA.  
File Name: Tech Bulletin_SQA_Blank Screen_12_OCT_10
Technical Bulletin‐ Blank Lower Operational Screen  
Applies to the following:  ALL SQA‐V (V, Vb, Ve, Vp and Vt) 
Issue date:  October 12, 2010  
 
 
 
 
www.mes-global.com
 
Note the alignment of
the red line on the cable
 
MB side   
 
Figure 1: Verify the flat cable is properly connected 
Figure 2: J5 connector on the main board
www.a-tech-global.com
        
 
 
LCD screen side 
J5
connector
 
 
Page 1 of 3