A.H. Systems SAS-542 Manuale operativo - Pagina 10
Sfoglia online o scarica il pdf Manuale operativo per Antenna A.H. Systems SAS-542. A.H. Systems SAS-542 13. Biconical antenna
Anche per A.H. Systems SAS-542: Manuale operativo (9 pagine)
A.H. Systems SAS-542 Biconical Antenna
CALCULATIONS
EMISSIONS TESTING
Individual calibration data for the log periodic antenna is supplied at appropriate
distances (3, and 10 meter) to comply with various emissions test requirements.
For emissions measurements, add antenna factor plus cable loss to receiver
reading in dBV to convert to field strength in dBV/meter.
Field Strength(dBuV/m) = SA(dBuV) + AF(dB/m) + cable loss (dB)
IMMUNITY TESTING
For Immunity measurements, the generated electric field strength can be
calculated by:
© A.H. Systems inc. –REV C
SA = Spectrum Analyzer or Receiver voltage reading
AF = Antenna Correction Factor
CL = Cable Loss in dB
FS = Approximate Field Strength in (V/m)
P = Power in watts
g = Numeric Gain
d = Distance in meters
10