A.H. Systems PAM-0118P Manuale operativo - Pagina 5
Sfoglia online o scarica il pdf Manuale operativo per Amplificatore A.H. Systems PAM-0118P. A.H. Systems PAM-0118P 13.
A.H. Systems Model PAM-0118P Pre-Amplifier
Most common causes of failure:
Excessive power applied to the RF input (such as ESD to the input
Improper connection sequence (sequence such as application of signals to
Excessive DC voltage applied
Storage near strong EMI or RFI fields
Loose connections causing intermittent events
An over voltage at the preamp input will damage the preamplifier. This type of
damage is without fault of A.H. Systems and will void the warranty.
To connect:
1. Apply DC Voltage
2. Connect output load
3. Apply RF input
To disconnect:
1. Remove RF input
2. Remove load
3. Remove DC voltage
RANGE OF ENVIRONMENTAL CONDITIONS
This equipment is designed to be safe under the following environmental
conditions:
While the equipment will not cause hazardous conditions over this
environmental range, performance may vary.
A.H. Systems inc. – May 2014
REV D
connector or antenna connected to the input connector)
the preamplifier before powering up)
Indoor use
Altitude: up to 2 km
Temperature: 5 C to 40 C
Maximum relative humidity: 80% for temperatures up to 31 C.
Decreasing linearly to 50% at 40 C
Pollution degree 2: Normally non-conductive with occasional
condensation.
DC supply voltage fluctuations not to exceed +/- 10% of the nominal
voltage.