A.H. Systems Injection Current Probe Series
ANTENNA FORMULAS AND CALCULATIONS
A specific Transfer Impedance Factor is associated with each frequency. This number is
to be subtracted from the receiver reading in order to convert to dBuA
dBA = dBV (from Receiver) - Transfer Impedance (dB) + cable Loss (dB)
EXAMPLE:
Assume that the frequency of interest is 80 KHz and we need to find the current at
this frequency. Connect the probe per figure 1
Frequency: 80 KHz
Transfer Impedance: -2.76 dB ohms
Cable Loss: 0.1dB
Receiver Reading: -33.0dBuV
dBA = dBV (from Receiver) - Transfer Impedance (dB) + cable Loss (dB)
dBA =
dBA = -30.34
When injecting current:
Injection current (dB) = Input current (dB) - Insertion loss (dB) - Cable loss
Maximum power rating : 100 Watts for 30 minutes
Do not handle injection probe while testing is in progress.
Hazardous voltages may be present.