AB Sciex 4000 QTRAP Manuale hardware - Pagina 8
Sfoglia online o scarica il pdf Manuale hardware per Apparecchiature di laboratorio AB Sciex 4000 QTRAP. AB Sciex 4000 QTRAP 18.
System Overview
Figure 1-2 Back view of the instrument
Component Descriptions
This section contains general descriptions of the components of the instrument.
Table 1-1 Component Descriptions
Component
I/O Panel
Gas and Vacuum Panel
Control Panel
Source Exhaust System
Vacuum Chamber
8
Turbo Pump
RF FEEDBACK
B
A
I/O Panel
Description
The external power supply, the IEEE-488 (GPIB) connections
to the computer, the AUX I/O to your peripherals and the
Sources connection are made through the I/O panel on the
right side of the chassis.
The vacuum lines to the external roughing pump are
connected through the gas and vacuum panel. The panel also
houses the gas supply connections and the external
connection for the source exhaust pump.
The control panel contains the main power switch, four fuses,
and two Indicator/Reset buttons.
Most ion sources produce both sample and solvent vapors.
These vapors are a potential hazard to the laboratory
environment. The source exhaust system is designed to safely
remove and allow for the appropriate handling of the ion
source exhaust products.
The vacuum chamber houses the quadrupole rod sets, most
of the ion optics, the collision cell, and the ion detector.
Turbo Pump
RF FEEDBACK
B
A
Gas and Vacuum Panel
Document Number: D1000092200 B