The Baclyser
neo S is a
-marked, sterile, point-of-use
®
shower filter
for use in high-risk areas in medical facilities for
1
the prevention of bacterial nosocomial infections caused by
Pseudomonas aeruginosa, Legionella pneumophila, Stenotro-
phomonas maltophilia or Staphylococcus aureus in immuno-
compromised and immunosuppressed patients. The effect is
achieved by interrupting the transmission pathway of these
germs through filtration of the water and protection against
retrograde contamination during use of the water taps through
hygienic design and antimicrobial protection functions. The
Baclyser
neo S is available in 3 different service lives: 2M, 3M
®
and 4M.

Technical data

Size: ø Shower head: 92 mm,
ø Handle: 41 mm, Length: 256 mm
Retention: 7 log steps
2
Membrane: Hollow fibre, pore size 0.2 μm
Operating time: 62, 93 or 124 days (2M,
3M or 4M) 
3
Flow rate: 15 l/min at 5 bar, 11 l/min
at 3 bar
Max. operating pressure: 5 bar 
Max. operating temperature: 60 °C
(at 70 °C ≤ 30 min over operating time)
Connection: G ½" thread
Chlorine resistance: ≤ 10 ppm 
Chlorine dioxide resistance: ≤ 0,8 mg/l
92 mm
256 mm
Sterile filtration: Path-
1
ogen retention ≥ 10
diminuta CFU/cm
membrane surface (= 7
log steps).
Measured in accordance
2
with the international
ASTM F838-20 standard
4
– Standard Test Method
for Determining Bacterial
Retention of Membrane
Filters Utilized for Liquid
Filtration.
Depending on water qual-
3
5
ity. If you use the Aqua
free filter change service,
the filter will be changed
every 8, 12 or 16 weeks.
The Baclyser
neo S
4
®
must not be connected
to low pressure units
(boiler).
Continuous dosage of ≤
5
10 ppm over operating
time; short-term (1h) high
dosage (400,000 ppm)
for chemical disinfection.