Allied Healthcare VACUTRON CONTINUOUS-INTERMITTENT Manuale di funzionamento e manutenzione - Pagina 9

Sfoglia online o scarica il pdf Manuale di funzionamento e manutenzione per Apparecchiature mediche Allied Healthcare VACUTRON CONTINUOUS-INTERMITTENT. Allied Healthcare VACUTRON CONTINUOUS-INTERMITTENT 16. Vacuum regulators for continuous or intermittent drainage

Allied Healthcare VACUTRON CONTINUOUS-INTERMITTENT Manuale di funzionamento e manutenzione
6.6

Flushing:

WARNING
6.6.1
Turn the mode selector switch to OFF. Attach two vacuum hoses as shown in Figure 8. The
collection bottle will trap the liquid and keep it out of the hospital vacuum system. Place the patient
connection of the Vacutron® regulator into a shallow pan containing isopropyl alcohol.
6.6.2
Turn the mode selector switch to the desired mode and turn the vacuum level knob (lower large
knob) clockwise as far as it will go for maximum flow. Draw approximately 100 ml of
isopropyl alcohol through the unit.
6.6.3
Remove the unit from the alcohol container and continue to operate the unit for at least 30
minutes with the patient connection of the regulator open to the atmosphere. This allows the
internal passages to dry out.
6.6.4
Test the unit in accordance with Section 2.0.
Figure 6
FLUSH THE UNIT IN EACH FUNCTIONAL MODE.
Figure 8
9