CPS TEMP-SEEKER TM360 Manuale d'uso - Pagina 7
Sfoglia online o scarica il pdf Manuale d'uso per Strumenti di misura CPS TEMP-SEEKER TM360. CPS TEMP-SEEKER TM360 20. Precision thermo-psychrometer
Anche per CPS TEMP-SEEKER TM360: Manuale operativo (13 pagine)
TMX3RH METHOD
The TMX3RH is a high accuracy silicon device
for the measurement of relative humidity and dry
bulb temperature. Both sensors are integrated
in one small silicon chip, factory calibrated and
conditioned to give accurate and fast response. The
TM360A uses this accurate data to compute the
wet bulb and the dew point temperatures using the
latest software algorithms.
Wet Bulb Temperature Accuracy (1 Atmosphere): +/-1.5% or +/- 1.0˚F (+/-0.5˚C) from
32˚F to 120˚F (0˚C to 50˚C) and 20% to 80% RH.
Dew Point Temperature Accuracy (1 Atmosphere): +/-1.5% +/-1.0˚F(+/-0.5˚C) from 32˚F
to 120˚F ( 0˚C to 50˚C) and 20% to 80% RH.
WET-SOCK METHOD
This method can be used to obtain relative humidity, dry/wet bulb and dew point measure-
ments using two standard TMX2G temperature probes and channels A and B. There are two
options for Step 4. Both options provide accurate readings and are merely a method of user
preference.
Step 1: Connect a TMX2G probe to Channel A and
another to Channel B.
Step 2: Press the RH key and then the MIN
key to display the minimum value of the RH.
Step 3: Slip a supplied 2" cotton sock* over the
end of the TMX2G probe connected to
Channel B and wet it with distilled water.
Step 4 (Option 1): Grip both probes approximately 18" from the ends in one hand.
Proceed to swing the probes in a circular motion through the air until the
display reading stabilizes. At that point you can stop swinging the probes.
Step 4 (Option 2): Place both probes in front of an air stream from air handler or
register. For best results the air should flow at about 9 feet per second (3 m.
second) over the wet sock ensuring that no radiant heat flows into the probe
such as direct sunlight or heat from machinery. The correct measurement is
obtained when the display stabilizes.
Step 5: Press the CLR key to erase the readings during the stabilization period and
display the current values. Pressing the RH,DB,WB or DP keys will display the
current value of the relative humidity, dry bulb, wet bulb, or dew point
temperatures.
* TMX3WB 2" cotton socks for RH measurement using wet sock method (6 pieces)
The computed Wet Bulb and Dew Point specifications
TMX2G w/
7
2" cotton sock