Results Flags (Partial list)
Capital letters indicate WBC related flags; lower-case letters indicate RBC
or PLT-related flags. For more details, consult your HM5 operator's manual.
Flag
Meaning
E
Empty sample
M,N
Linearity error (WBC)
- cell concentration
too high
C,Q
WBC clogging
L
Potential lyse resistant
RBCs in canine samples
X
EOX blank high
Y
EOS clogging
Z
Noise in EOS measurement Check Lyse2 volume and
b
RBC blank is high
c
RBC/PLT clogging
p
PLT blank is high
Solution
Lower needle height
Prime lyse and repeat blank. View
manual smear to confirm high WBC
Run 3-5 self cleanings and enzymatic
cleaning
Re-run sample at higher
lyse volume
Check that Lyse2 is room tempera-
ture
Run 3 automated self cleanings and
repeat analysis
prime Lyse2 if needed
Clean wash head; run self cleaning;
bleach clean analyzer/tubes; change
reagents if necessary
Ensure that reagents are room
temperature run self cleanings and
enzymatic cleanings
Clean wash head; repeat blank 2-3x;
run 5 cycle cleaning; run enzymatic
cleaning; bleach clean analyzer/tub-
ing; change reagent pack