Abaxis VetScan HM5 Manuale di riferimento rapido - Pagina 11
Sfoglia online o scarica il pdf Manuale di riferimento rapido per Strumenti di misura Abaxis VetScan HM5. Abaxis VetScan HM5 19. Hematology analyzer
Anche per Abaxis VetScan HM5: Manuale di riferimento rapido (17 pagine), Manuale di installazione (1 pagine), Istruzioni per l'installazione (2 pagine)
This cleaning is meant to be used as an emergency procedure when a high HCT trend is
noticed: and it is not part of the regularly scheduled maintenance.
Please ensure that your HM5 is on software version 2.3 or higher.
1. Deep Cleaning. Go to Maintenance> Cleaning> Deep Cleaning.
Deep Cleaning
Cleaning & Maintenance Procedures
1a. Remove the cap
and place the VetScan
HemaClean tube into
the tube holder in the front
of the analyzer.
Touch Next to start
the cleaning. It takes
approximately 20 minutes
to complete.