STEP 2: CHANGING BATTERY PACKS
Each day, preferably at the same time to ensure
treatment is continued without interruption, patients
should do the following:
A. Slide down the battery door on the back of the
stimulator and remove the depleted battery
(Figure 7).
B. Place the depleted battery pack into the battery
charger for charging (Figure 8). A solid orange light
on the charger will illuminate, indicating the battery
is charging. If no light appears on the charging
cradle an error is indicated. If this occurs, try
removing the battery pack from the charger and
reinserting it. If the orange light does not appear
contact Biomet.
C. Once the charger's orange light turns off and
a solid green light appears, the battery pack is fully
charged. Remove the battery pack from the battery
charger (Figure 9) and place the fully charged battery
®
pack into the Biomet
SpinalPak
®
Stimulator in order
to commence treatment.
D. There should always be one battery in the charger
and one battery installed in the stimulator at all times,
ensuring a fully charged battery every 24 hours as
recommended.
NOTE: Do not be concerned if the battery packs are inadvertently charged more than once or kept
on the charger for a long period of time. The batteries cannot be overcharged. If the battery pack
is in the battery pack charger and the battery pack is fully charged, the charger will terminate the
recharging process. The charger having no orange light when the charging is complete will
indicate this termination. Additional replacement battery packs are available by contacting
Biomet.
Button Functions
ALARM ON/OFF BUTTON
®
The Biomet
SpinalPak
located below the LCD display enables or disables the audible alarm. During an alarm condition,
depressing the button quickly (0.5 seconds) will temporarily disable the audible alarm. Depressing the
button for a longer period of time (3 seconds) will toggle the audible alarm between enabled and dis-
abled. Patients should be advised to leave the audible alarm enabled as frequently as possible in order
to assure the fully prescribed treatment. A speaker symbol will be indicated on the LCD display when
Figure 7
the alarm is enabled.
LCD Symbol Descriptions and Instructions
The alarm defaults to audible alarm. Press the button below the display on the front of the stimulator
to silence the alarm. After silenced, the light will continue to flash and the display will indicate the
alarm condition.
Symbol/TEXT
Figure 8
Figure 9
8
9
®
Stimulator is activated as soon as a charged battery is inserted. The button
Condition
Instructions
Treating
Continue use.
Audible alarm
If beeping, depress the button
briefly to silence the alarm. Depress the
button approximately 3 seconds to
engage or disengage the audible alarm.
Low battery
Insert a charged battery pack.
Disconnection
Confirm that each electrode is properly
applied on the skin. See the electrode
pouch for instructions. Confirm that the
lead wire is attached properly. Replace
the lead wire if necessary.
System error
Error in the stimulator –
Contact Biomet for assistance.
Stimulator is
Stimulator will not treat until USB
connected to
cable is disconnected
a PC
End of operation
Contact Biomet