5.0 Installation in the Laboratory

Use the supplied power cable to connect the mixer/shaker to the mains supply.
The appliance connector socket is at the rear.
The mixer/shaker is protected by bipolar fuses. The fuses are located in the housing of the
appliance connector socket and can be reached from the outside. The fuse rating is printed on
the plate.
For a description of how to operate your mixer/shaker, turn to the section "7.0 Operating the
mixers/shakers" and look up your particular model variant.
When working with low shaking frequencies, you can place flat vessels, e.g. Petri dishes, on
the platform without an additional fixture.
This mixer/shaker is fitted as standard with a EURO plug (DIN 49441 CEE 7/VII 10/16 A
250 V). For North America it comes with a US standard plug (NEMA Pub. No. WDI-1961
ASA C 73.1 - 1951 page 8 15A 125 V).
To connect the mixer/shaker to a different mains system, you must use an authorized adapter
or you must have an electrician replace the supplied plug with a suitable 3-pin plug
authorized for your mains system. In its delivered condition the mixer/shaker is earthed! If
you replace the original plug, make sure that the protective earth conductor of your
mixer/shaker is connected to the new plug.
Note the following colour codes if you ever use a different power cable or make any changes
to the power cable supplied:
GREEN/YELLOW = protective earth conductor (PE)
BLUE
BROWN
IN THE UNITED KINGDOM IT IS COMPULSORY TO USE AN APPROVED
STANDARD 13 A PLUG IN CONFORMANCE WITH BS 1363 (3-PIN PLUG).
= phase (N)
= phase (L)
5