HTC CM-2007 Manuale - Pagina 3
Sfoglia online o scarica il pdf Manuale per Strumenti di misura HTC CM-2007. HTC CM-2007 5. Digital clamp meter
prevent the electric shock.
Function choosing switch: Used to select the function, range or turn
on or turn off the clamp meter.
Function
Description
ACA measurement.
ACV, D CV m easur e ment . P ress " SELECT" t o s wi t ch t he m easur e ment m ode. P r e ss " Hz/DUTY" t o s wi t c h t o F r e quency/Dut y c ycl e m easurement m ode
NCV
Non-contact voltage test.
Press "SELECT" to switch to Diode, Continuity,
Ω
Capacitance and Resistance measurement mode.
Hz
Frequency measurement.
Temperature measurement. Press "SELECT" button to switch
℃ ℉
/
to ℃ or ℉.
Function key.
LIGHT button: Backlight and Clamp light control key. Press
LIGHT button more than 2 seconds, backlight and clamp light will be
on. Press this button again, the backlight and clamp light will be off.
The backlight will auto turn off after 5 seconds if not pressing it again.
SELECT button: Function selection key. Press this button to select the
measurement mode: Under "
" mode, choose DC or AC. Under
Ω
"
" mode, choose "
" "
" "
", "
HOLD button: Press "HOLD" button, the current testing value will
be kept displayed on the LCD, and press this button again to cancel this
function. When this function works, user cannot do any new test.
REL button: Press this button to clear the readings and enter into the
Relative measurement mode and a "REL" symbol will be displayed on
the LCD. Press this button again to cancel this function.
LCD: Displaying measurement value and unit.
.
V/Ω input terminal: Measurement inputs positive terminal (red
test lead).
COM input terminal: Measurement inputs negative terminal (black
test lead).
Carrying belt.
4.2 ACA MEASUREMENT
(1) Turn the rotary switch to
Ω"
measurement.
range.
(2) Open the clamp jaw and grip one
wire and get the reading directly.
To get the most precise reading, should user put the wire in the
middle of the close jaw
Note:
1.
Firstly users should select the highest range, if users are
not sure about the range of current under test, and then select the
proper range based on displaying value.
No.
Symbol
Description
-
Indicates negative readings.
AC
AC measurement.
DC
DC measurement.
Auto
Auto range.
Null.
Diode test.
Continuity beeper is on.
HOLD
Data hold
REL
Relative measurement.
Low battery indication.
hFE
Null.
℃,℉
℉ null.
NCV
Non-contact voltage testing.
nF,uF
nF,uF
○
11
MΩ,kΩ,Ω
MΩ,kΩ,Ω
mV, V
mV, V
uA, mA, A
uA, mA, A
Hz,kHz,MHz
Hz,kHz,MHz
(refer to the picture 3).
2.
DO NOT input the current higher than the limit, or it
would damage the meter.
4.3
ACV MEASUREMENT
(1)
Turn the function swift to
range.
(2)
Insert the black test lead into "COM"
terminal, and the red one into "V/Ω" terminal.
(3)
Connect the test leads into the circuit and get
the measurement value from the LCD. (refer to
picture 4)
Note:
1. DO NOT input the voltage over the limit, or it would damage
the meter.
2. Be careful while measuring a high voltage circuit. Do not touch
the high voltage circuit.
4.4 DCV MEASUREMENT
(1) Turn the function swift to DCV range.
(2) Insert the black test lead into "COM" terminal, and
the red one into "V/Ω" terminal.
(3) Connect test leads to the test point. LCD will display
polarity and voltage of the test point connected by the
red test lead. (refer to picture 5)
Note:
1. DO NOT input the voltage over the limit, or it would damage
the meter.
2. Be careful while measuring a high voltage circuit. Do not touch
the high voltage circuit.
4.5 NCV MEASUREMENT
NCV Static electricity or other sources of energy
may randomly trigger the sensor. The result is only
for reference.
4.5.1 Turn the rotary switch to NCV position.
4.5.2 Hold the top of the meter very close to the
voltage
source
under
test.
4.5.3 If voltage is present, the NCV LED will flicker rapidly or
intermittently
NOTE:
1. Even no indication when NCV test, there might still voltage be.
(picture 4)