Dip Switch Setup
To set-up the ET8 recorder for your specific application,
you might need to change some of the "Dip Switches". The
Dip Switches are located on the back of the unit.
Use a pen or small screw driver to flip the switches.
Remember to install the correct chart to match the
corresponding Dip Switch setting.
Slide toggle up for "on" position
Slide toggle up for "on" position
Dip Switch 1
1
2
5–8 are
not active on
this recorder.
Slide toggle down for "off" position
ET8 Dip Switch Setting Up=On Down=Off
Switches
1 2 3 4 5(ch2) 6(ch1)
1–8
D
7-Day
U
24-Hour
D D D
0.0 to 100.0
D D U
0 to 1000
D U D
0.0 to 14.0
D U U
0.0 to 150.0
U D D
0.0 to 200.0
U D U
0.0 to 300.0
U U D
-10.0 to 10.0
U U U
-150.0 to 250.0
0.0 to 5.0 volts
0.0 to 20.0 ma
1.0 to 5.0 volts
4.0 to 20.0 ma
Remember to install the correct chart to match corresponding switch setting
DICKSON • Addison, IL • www.DicksonData.com
Make sure switches 5,6,7 & 8 are set correctly to match the
type of transducer you are using and that the transducer is
plugged into the correct jack.
Recording Time
The ET8 recorder has two different recording time options
24-hour
1 up
7-day
1 down
www.DicksonData.com • 1-800-323-2448 • Fax 1-800-676-0498
3
4
5
6
7
8
7(ch2)/8(ch1
D
D
D
U
U
D
U
U
Recording Range
Range
0.0 to 100.0
0.0 to 1000.0
0.0 to 14.0
0.0 to 150.0
0.0 to 200.0
0.0 to 300.0
-10.0 to 10.0
-150.0 to 250.0
Channel Input
(Channel 1= Red Pen; Channel 2=Blue Pen)
Channel Input
2
0.0 to 5.0V
2
0.0 to 20.0 ma
2
1.0 to 5.0V
2
4.0 to 20.0 ma
1
0.0 to 5.0V
1
0.0 to 20.0 ma
1
1.0 to 5.0V
1
4.0 to 20.0 ma
Power
The ET8 operates on AC power with an optional 9V battery
backup. Battery backup will last approximately 3 days.
Alarm and relays will not operate under battery only power.
LED Indicators
• AC power with battery backup–Solid green
• AC power with low battery or no battery–Blinks red
• Battery only–Blinks green
• Battery only (low battery) –Solid red
ET8
Universal Input
Chart Recorder
2
3
4
D
D
D
D
D
U
D
U
D
D
U
U
U
D
D
U
D
U
U
U
D
U
U
U
5
6
7
8
D
D
D
U
U
D
U
U
D
D
U
U
D
U
D
U