Alpha MED Scientific MED64 MED-C11 Manuale del prodotto - Pagina 14
Sfoglia online o scarica il pdf Manuale del prodotto per Cavi e connettori Alpha MED Scientific MED64 MED-C11. Alpha MED Scientific MED64 MED-C11 20. Med multi-well connector
4. Troubleshoot
4. Troubleshoot
When noise appears during your experiments, it is important to identify the source of the noise. Likely
reasons are:
• Damaged electrodes of the MED Probes
• Uncleaned and/or damaged contact pins in the Connector.
• Loosely connected the cables.
• Improper grounding.
• Equipment that generates and electric or magnetitic field (e.g. power supply, peristaltic pump,
heater).
• Environment
Poor contacts caused by damaged or uncleaned contact pins can cause problems such as noises or/and
faulty recording. Identify this problem by using the Test board following this procedure:
Figure 14. The Test board installed in the MED Multi-well Connector.
Trouble shooting flow chart
10
Product Manual - MED Multi-well Connector (MED-C11)