Olympus DP70 Manuale d'uso - Pagina 30
Sfoglia online o scarica il pdf Manuale d'uso per Fotocamera digitale Olympus DP70. Olympus DP70 38. Application software
Help with problems
Most of the problems concerning image quality can be solved by changing the cam-
era settings. To open the
located in the
dow either on the right hand side of your screen, or, if you chose the
Style
Should you not be able to solve your problem with the instructions shown here,
please contact the Olympus customer service.
Problem
Your computer doesn't start.
"Live" and "Snap" are inac-
tive.
No live-image can be seen.
With an acquired image, the
message "Error making a
snapshot" appears.
The live-image jerks.
The image looks too dark.
Camera Settings
Camera Control
when you installed your program, at the bottom of your screen.
Possible cause
The PCI board has not been
properly inserted in your PC's
PCI slot.
The connection between the
camera and your PC is faulty.
The device drivers have not been
correctly installed.
A false camera has been se-
lected in the
Camera Control
picklist located below the
and
Snap
buttons.
The lamp on your microscope is
not switched on.
The camera isn't in the light path.
Either the illumination has been
falsely set, or the sample falsely
positioned.
The connection between the
camera and your PC is faulty.
The diaphragm is not wide
enough open.
dialog box, click the
tool window. You will find the
Remedy
Switch your PC off, pull out the
power plug, then insert the PCI
board firmly in the PCI slot.
Plug the power plug back in, and
switch your PC on.
Switch your PC off, pull out the
power plug, then insert the plugs
on the connector cable firmly into
the PCI board and the camera.
Plug the power plug back in, and
switch your PC on.
Check the camera's installation.
Select the correct camera from
this picklist.
Live
Switch the lamp on.
Set the light path so that the light
falls on the camera.
Check the illumination and the
sample.
Switch your PC off, pull out the
power plug, then insert the plugs
on the connector cable firmly into
the PCI board and the camera.
Plug the power plug back in, and
switch your PC on.
Select a lower resolution.
Open the diaphragm on the mi-
croscope wider.
Help with problems
Settings
button
Camera Control
tool win-
DP-Controller
Page
30
(User's
Manual)
(User's
Manual)