Olympus BH2-CH Substage Montaggio - Pagina 9

Sfoglia online o scarica il pdf Montaggio per Microscope Olympus BH2-CH Substage. Olympus BH2-CH Substage 26.

Olympus BH2-CH Substage Montaggio
Note the location of any small, brass shim washers that
may be present beneath any of the slip plates so that
they can be returned to their proper place during
reassembly (see
Figure
10).
Figure 10 – Look for shim washers under the slip plates
Carefully remove any brass shim washers that are
present beneath the slip plates (see
Figure 11 – Remove any brass shim washers

Remove the Centering Yoke from the Support Yoke

Place your thumb in position on the centering yoke as
shown in
Figure
12, and press in on the centering yoke
to depress the centering plunger.
Figure 12 – Depress the centering plunger
Complete Teardown, Cleaning/Lubrication, and Reassembly of the Olympus BH2-CH Substage Assembly
Figure
11).
While holding the centering yoke such that the
centering plunger is depressed, separate the centering
yoke from the support yoke (see
Figure 13 – Separate centering yoke from support yoke
The various components of the support yoke (which is
part of the condenser slide) are shown in
Figure 14 – The support yoke on the condenser slide
The centering yoke is shown in
Figure 15 – The centering yoke

Remove the Condenser-Height Rack

Use a suitable JIS screwdriver to remove the two
M2.6x6 JIS screws and washers securing the condenser-
height rack to the dovetail slide (see
Figure
13).
Figure
14.
Figure
15.
Figure
16).
Revision 1
Page 9 of 26