8-2 Detailed Assembly Procedures
1
3
2
4
6
5
25
Fig. 12
Fig. 13
Fig. 14
Fig. 15
2
Attaching the Fluorescence Mirror Units
1. Hold the left and right sides of the cover @ on the front of the illuminator
and remove the cover by pulling it toward you.
2. Rotate the turret 3 so that the turret position 2 you want to mount the
fluorescence mirror unit comes on the front.
3. Pull out the dummy mirror unit | and push in the fluorescence mirror
unit until it clicks.
4. Push the inserted fluorescence mirror unit from both sides alternately.
Otherwise, the fluorescence mirror unit may be mounted being tilted.
} Two more fluorescence mirror units can be mounted in the turret positions
on the left and right in the current condition.
} Write down the turret position numbers and fluorescent mirror unit names
for use when you insert the indication sheets later.
5. Mount other necessary fluorescence mirror units in the same way as
above. Do not remove the dummy mirror units from the turret positions
you do not want to mount fluorescence mirror units.
6. Attach the cover @ in the original position.
7. Insert the indicator sheet 5 for the fluorescence mirror unit mounted in
each turret position 2 (Fig. 13) under the film of the corresponding mirror
unit inscription pocket 6.
(Figs. 12 to 15)
(Fig. 14)